Abstract
-
Objective
To investigate the reliability and validity of a new method for isometric back extensor strength measurement using a portable dynamometer.
-
Methods
A chair equipped with a small portable dynamometer was designed (Power Track II Commander Muscle Tester). A total of 15 men (mean age, 34.8±7.5 years) and 15 women (mean age, 33.1±5.5 years) with no current back problems or previous history of back surgery were recruited. Subjects were asked to push the back of the chair while seated, and their isometric back extensor strength was measured by the portable dynamometer. Test-retest reliability was assessed with intraclass correlation coefficient (ICC). For the validity assessment, isometric back extensor strength of all subjects was measured by a widely used physical performance evaluation instrument, BTE PrimusRS system. The limit of agreement (LoA) from the Bland-Altman plot was evaluated between two methods.
-
Results
The test-retest reliability was excellent (ICC=0.82; 95% confidence interval, 0.65–0.91). The Bland-Altman plots demonstrated acceptable agreement between the two methods: the lower 95% LoA was −63.1 N and the upper 95% LoA was 61.1 N.
-
Conclusion
This study shows that isometric back extensor strength measurement using a portable dynamometer has good reliability and validity.
-
Keywords: Back muscles, Isometric contraction, Muscle strength dynamometer, Reproducibility of results
INTRODUCTION
Back pain is a common musculoskeletal condition that affects both men and women of different age groups. Due to the multi-factorial nature, it is difficult to identify exact causes of back pain, and treatment is a difficult and time-consuming process. Recently, back rehabilitation programs, which are non-surgical, multi-disciplinary treatments consisting of exercise and physical therapy, have gained popularity for prevention of spinal injury and back pain, and resuming performance of daily activities [
123456].
The key interventions involved in back rehabilitation programs are strengthening and retraining of back muscles [
78910]. Weakness of back muscles has been identified in many back pain cases [
6711]. Back muscle strength is, therefore, the main target of intervention in back rehabilitation, and evaluation of back muscle strength is crucial [
2369]. Previous studies utilized an isokinetic dynamometer machine for evaluating back muscle strength in the isokinetic or isometric test modes [
6710]. However, an isokinetic dynamometer machine is neither readily available nor easily movable, and requires sophisticated maintenance efforts and well-trained operators [
78111213].
A portable dynamometer is a small hand-held device that can quantify the muscle strength of extremities in manual muscle testing. The portable dynamometer has a compression force sensor and a control/recording unit. The force sensor is small enough to be mounted on an examiner's hand while resisting against an examinee's action (i.e., hand-held dynamometer). The control/recording unit displays the maximal compression force. To measure the strength using the portable dynamometer, the examiner must use his/her own strength to fully resist the action performed by the subjects. The hand-held dynamometer has not been applied to the measurement of trunk muscle strength because it is difficult for most examiners to resist the examinee's trunk action. The authors have designed a special chair with a seatback that can hold a force sensor unit and resist isometric back extension movement (
Fig. 1).
The aim was to assess the reliability of this new method, which combined a portable dynamometer with a chair for isometric back strength evaluation, and compare its strength data with that of the conventional dynamometer system for a validity analysis.
MATERIALS AND METHODS
The Institutional Review Board of Kangwon National University Hospital approved this study protocol (2014-01-007-004), and informed consent was obtained from all participants.
Study design
This was a prospective, comparative, test-retest reliability and validation study. For test-retest reliability analysis, different day values of back strength data measured by portable dynamometer were used. We used a commercially available portable dynamometer (Power Track II Commander Muscle Tester; JTECH Medical, Midvale, UT, USA) originally designed for manual muscle strength testing of extremities. This machine is able to record multiple peak compression forces, and the highest force generated by a subject is recorded automatically at each attempt.
For test-retest reliability, a second measurement was assessed on a different day. On the day of the second measurement, the isometric back extensor strength was also measured using physical performance evaluation instrument, PrimusRS (BTE technologies Inc., Hanover, MD, USA) for the validation study. The PrimusRS is an isokinetic dynamometer machine able to assess isometric, isotonic, and isokinetic muscle forces. The PrimusRS can record accurately up to 1800 in. lbs. of isometric force.
Participants
Fifteen men (mean age, 34.8±7.5 years) and 15 women (mean age, 33.1±5.5 years) aged 19–60 years were included. Subjects were interviewed by a physician and only those who had no current back problems or previous history of back surgery were enrolled.
Device design
We designed a special chair for measuring isometric back extensor strength using the portable dynamometer (
Fig. 1). It is a typical chair with a seat and seatback, but in the center of the seatback, there is a vertical linear hole for holding a sensor unit of the portable dynamometer and a square back plate for stable transmission of pressure generated by the back extensors of a seated subject to the sensor unit. The sensor unit and back plate position is modifiable along the seatback hole to the intended height because a ruler was attached to the seatback. In the lower part of the seat, there is a seat belt for safety and body stabilization. The seat was high enough to lift the feet of subjects off the ground to minimize any force exerted by legs.
Measurement of isometric back extensor strength
Initially, the subject was seated in the chair and an examiner checked the subject's height of the T7 vertebral spinous process. Then, the pelvis was fixed by fastening the seat belt. The sensor unit of the portable dynamometer was positioned in the vertical seatback hole with the center of sensor unit at the T7 spinous process level. The square back plate was positioned between the subject and seatback. The subject placed arms on chest folded and kept shoulders relaxed. With the starting signal, the subject was asked to push the square back plate with his/her back as far as he/she could for 3 seconds. This process was repeated 5 times, and the mean value was calculated only with 3 median values, excluding the maximum and minimum values. For second measurement, the entire process was repeated for each subject on different days. Finally, two processed values for each participant were acquired for test-retest reliability assessment.
On the same day of the second measurement using the portable dynamometer, the isometric back extensor strength was measured using the PrimusRS system, an isokinetic dynamometer machine (
Fig. 2). Between two measurements using the portable dynamometer and PrimusRS system, subjects rested for more than 2 hours. The subjects were seated in the same chair described for the portable dynamometer evaluation and wore an evaluation harness anchored to the PrimusRS with a cable. The height of the anchoring cable was modified to the T7 spinous process level with the subjects seated in the chair. They were instructed to maintain an upright posture without leaning on the seatback and shoulders relaxed. With the starting signal, they were asked to push the square back plate with their back as far as they could for 3 seconds with isometric back extension. When subjects did the isometric back extension, the cable anchored to the chest harness was pulled and isometric back extension strength was measured. This process was repeated 5 times, and the mean of 3 median values excluding the maximum and minimum values were calculated. The results were compared with the second measurement data obtained using the portable dynamometer for the validation study.
Statistics
We calculated the intraclass correlation coefficient (ICC) for assessing test-retest reliability. For validation assessment, we used Pearson correlation coefficient and Bland-Altman plotting and 95% limits of agreement (95% LoA) were computed. All statistical analyses were performed using SPSS ver. 20 (IBM, Armonk, NY, USA).
RESULTS
All demographic data, as well as the mean isometric back extensor strength values at different test times, are shown in
Table 1. The isometric back extensor strength values evaluated by first and second isometric tests using the hand-held dynamometer were 204.9±52.6 N and 205.9±49.5 N, respectively. The isometric strength measured by PrimusRS was 212.6±51.6 N.
The test-retest reliability of the measured strength showed good correlation. The ICC of two tests using the hand-held dynamometer was 0.82 (95% confidence interval, 0.65–0.91) (
Table 2).
In the correlation study comparing the PrimusRS system data and our new method, Pearson r was 0.65 (
Fig. 3). Bland-Altman plots demonstrated acceptable agreement (
Fig. 4) and the 95% LoA for the PrimusRS method and our new method was −63.1 N to 61.1 N.
DISCUSSION
To assess the test-retest reliability of our new method for isometric back extensor strength evaluation, ICC was utilized. The ICC measures the relative similarity of quantities that share the same observational units of a sampling and/or measurement process [
14]. The ICC is considered acceptable at 0.80 or higher. The ICC of our new method demonstrated 0.82, which indicates non-inferior reliability to that of previous studies using the isokinetic dynamometer machine [
812].
In validation analysis comparing established and unestablished assessment technology, the Bland-Altman plot is considered the most appropriate method for measuring agreement [
1415]. More classical measures such as correlation coefficient (r) and linear regression are inappropriate and often misleading because their primary function is to analyze the strength of the relationship and not the agreement between two variables [
15]. In this study, Bland-Altman plots demonstrated that there was an acceptable agreement between the back extensor strength data obtained with the PrimusRS and the data obtained using our new device. The procedure described by Bland and Altman was developed to examine the agreement between two different techniques of quantifying variables, and not necessarily the reliability. The use of the Bland-Altman plot and the limits of agreement as an index of reliability is typically discouraged [
14].
Our target structure of strength measurement was the back extensor muscle group. This muscle group mainly consists of multifidus and erector spinae muscles. In muscle fiber type analysis in normal population, the type 1 (slow twitch) muscle fiber was predominant in lumbar multifidus and erector spinae muscles [
16], and this may be a reflection of the functional role of the back extensor group as a core stabilizer requiring a high level of endurance. The strength of back extensor muscles decreases in older age as well as female gender, and the rate of decline in strength with age was greater in the aged group and two times higher in women [
5]. The incompetence of back extensor muscle is reported to be associated with back pain in both young and old populations [
71317]. It has been documented that in patients with chronic back pain, the cross-sectional area of the back muscle group on a CT scan was smaller than that of the healthy control group, while the size of the paraspinal muscles was correlated with the isometric back extension strength.
Back muscles protect pain-sensitive structures, such as vertebral columns, intervertebral discs, and facet joints, from excessive stress and loading, therefore the compromising of back muscles may result in injury to these passive structures and induce back pain [
410]. Due to their importance, the strengthening of back muscles may be essential to treat and prevent lower back problems [
24]. In recent years, several systematic reviews have demonstrated that rehabilitative exercise therapy is an effective approach in the treatment of chronic and recurrent back pain [
181920]. Various types of rehabilitation exercises including endurance training, resistive weight training, and stabilization exercises are known to have a beneficial effect in back pain patients, but there is a lack of evidence, as well as a definitive exercise protocol, for most of these exercise types.
We used a seat belt for both pelvic stabilization and hip movement minimization in the sitting position. Hip extensor muscles, such as the glutei and hamstrings, can play a role in back extension which can affect measurement accuracy. To minimize the impact of hip extensors during trunk extension measurement, pelvic stabilization is important [
2122]. Another study suggested that combining pelvic stabilization and the semi-sitting position with a hip flexion angle of 45° would be helpful in increasing the specificity of back extensor exercises [
23]. We focused on pelvic stabilization throughout the experiment, however difficulties in device design did not allow for the adaptation of the semi-sitting test position.
Strength testing using the isokinetic dynamometer is the standard quantitative assessment used for evaluation of trunk muscles [
24], however the isometric testing method should also be considered for assessing muscle strength [
2526]. In isometric muscle contraction, the length of the muscle remains constant and the angular velocity is zero. Almost every isokinetic dynamometer device can be used for isometric testing by simply selecting an angular velocity of zero [
24]. Joint angles, however, should be selected carefully in an isokinetic device due to significant force variations throughout the joint's range of motion [
24]. It is not clear which of the two, isokinetic or isometric strength measurements, is superior, especially in terms of reliability. Generally, it is believed that isokinetic testing is more responsive to back problems due to the considerable effect of joint movement on pain provocation [
817]. In previous studies on patients with back pain, isokinetic testing produced more varying results in comparison with isometric testing, and repetition of muscle strength evaluation showed significantly increased results, probably due to the ‘learning effect’ [
17]. We demonstrated good reliability of isometric trunk muscle strength assessment using our new device, and the learning effect was negligible.
Another merit of isometric back muscle testing is supported by evidence provided by a long-term outcome report. A 10-year follow up study on postmenopausal women by Sinaki et al. [
26] demonstrated that isometric back strength is closely related to the prevention of vertebral compression fracture. The new isometric testing method discussed here has the added benefits of a simple design and a device that is lightweight, easier to operate and can be transported by one person.
The subjects of this study were healthy volunteers with no current back problems or previous history of back surgery. The reliability could be lower in patients with back pain, although previous study showed comparable reliability of isokinetic measurements between normal and back pain subjects [
8]. There is a need for further investigation to assess reliability in the very old population and in those with current back pain.
Compared to the conventional methods, the advantage of the device we designed is high portability and ease of use. This equipment can be transported by one or two people and installed within 5 minutes. It is an ideal evaluation tool for group rehabilitation program or field study outside hospital in the local community.
In conclusion, we developed a simple method for isometric back extensor strength measurement using a specially designed chair and portable hand-held dynamometer. Our method has unmatched simplicity and mobility over the conventional dynamometer-based method, and it demonstrated good reliability and acceptable agreement with measurement data obtained using the isokinetic dynamometer machine in the healthy population.
ACKNOWLEDGMENTS
ACKNOWLEDGMENTS
This study was supported by 2015 Research Grant from Kangwon National University.
CONFLICT OF INTEREST:
CONFLICT OF INTEREST: No potential conflict of interest relevant to this article was reported.
REFERENCES
- 1. Little P, Smith L, Cantrell T, Chapman J, Langridge J, Pickering R. General practitioners' management of acute back pain: a survey of reported practice compared with clinical guidelines. BMJ 1996;312:485-488.
- 2. Chok B, Lee R, Latimer J, Tan SB. Endurance training of the trunk extensor muscles in people with subacute low back pain. Phys Ther 1999;79:1032-1042.
- 3. Kaser L, Mannion AF, Rhyner A, Weber E, Dvorak J, Muntener M. Active therapy for chronic low back pain: part 2. Effects on paraspinal muscle cross-sectional area, fiber type size, and distribution. Spine (Phila Pa 1976) 2001;26:909-919.
- 4. Moon HJ, Choi KH, Kim DH, Kim HJ, Cho YK, Lee KH, et al. Effect of lumbar stabilization and dynamic lumbar strengthening exercises in patients with chronic low back pain. Ann Rehabil Med 2013;37:110-117.
- 5. Singh DK, Bailey M, Lee R. Decline in lumbar extensor muscle strength the older adults: correlation with age, gender and spine morphology. BMC Musculoskelet Disord 2013;14:215.
- 6. Cho KH, Beom JW, Lee TS, Lim JH, Lee TH, Yuk JH. Trunk muscles strength as a risk factor for nonspecific low back pain: a pilot study. Ann Rehabil Med 2014;38:234-240.
- 7. Bayramoglu M, Akman MN, Kilinc S, Cetin N, Yavuz N, Ozker R. Isokinetic measurement of trunk muscle strength in women with chronic low-back pain. Am J Phys Med Rehabil 2001;80:650-655.
- 8. Keller A, Hellesnes J, Brox JI. Reliability of the isokinetic trunk extensor test, Biering-Sørensen test, and Astrand bicycle test: assessment of intraclass correlation coefficient and critical difference in patients with chronic low back pain and healthy individuals. Spine (Phila Pa 1976) 2001;26:771-777.
- 9. Hanada EY, Hubley-Kozey CL, McKeon MD, Gordon SA. The feasibility of measuring the activation of the trunk muscles in healthy older adults during trunk stability exercises. BMC Geriatr 2008;8:33.
- 10. Lee HJ, Lim WH, Park JW, Kwon BS, Ryu KH, Lee JH, et al. The relationship between cross sectional area and strength of back muscles in patients with chronic low back pain. Ann Rehabil Med 2012;36:173-181.
- 11. Bernard JC, Boudokhane S, Pujol A, Chaleat-Valayer E, Le Blay G, Deceuninck J. Isokinetic trunk muscle performance in pre-teens and teens with and without back pain. Ann Phys Rehabil Med 2014;57:38-54.
- 12. Hutten MM, Hermens HJ. Reliability of lumbar dynamometry measurements in patients with chronic low back pain with test-retest measurements on different days. Eur Spine J 1997;6:54-62.
- 13. Yahia A, Ghroubi S, Kharrat O, Jribi S, Elleuch M, Elleuch MH. A study of isokinetic trunk and knee muscle strength in patients with chronic sciatica. Ann Phys Rehabil Med 2010;53:239-244. 244-249.
- 14. Weir JP. Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res 2005;19:231-240.
- 15. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307-310.
- 16. Rantanen J, Rissanen A, Kalimo H. Lumbar muscle fiber size and type distribution in normal subjects. Eur Spine J 1994;3:331-335.
- 17. Gruther W, Wick F, Paul B, Leitner C, Posch M, Matzner M, et al. Diagnostic accuracy and reliability of muscle strength and endurance measurements in patients with chronic low back pain. J Rehabil Med 2009;41:613-619.
- 18. Pengel HM, Maher CG, Refshauge KM. Systematic review of conservative interventions for subacute low back pain. Clin Rehabil 2002;16:811-820.
- 19. Mayer J, Mooney V, Dagenais S. Evidence-informed management of chronic low back pain with lumbar extensor strengthening exercises. Spine J 2008;8:96-113.
- 20. Scharrer M, Ebenbichler G, Pieber K, Crevenna R, Gruther W, Zorn C, et al. A systematic review on the effectiveness of medical training therapy for subacute and chronic low back pain. Eur J Phys Rehabil Med 2012;48:361-370.
- 21. Graves JE, Webb DC, Pollock ML, Matkozich J, Leggett SH, Carpenter DM, et al. Pelvic stabilization during resistance training: its effect on the development of lumbar extension strength. Arch Phys Med Rehabil 1994;75:210-215.
- 22. da Silva RA, Lariviere C, Arsenault AB, Nadeau S, Plamondon A. Pelvic stabilization and semisitting position increase the specificity of back exercises. Med Sci Sports Exerc 2009;41:435-443.
- 23. da Silva RA, Lariviere C, Arsenault AB, Nadeau S, Plamondon A. Effect of pelvic stabilization and hip position on trunk extensor activity during back extension exercises on a Roman chair. J Rehabil Med 2009;41:136-142.
- 24. Maud PJ, Foster C. Physiological assessment of human fitness. 2nd ed. Champaign, IL: Human Kinetics; 2006. p. 131-138.
- 25. Limburg PJ, Sinaki M, Rogers JW, Caskey PE, Pierskalla BK. A useful technique for measurement of back strength in osteoporotic and elderly patients. Mayo Clin Proc 1991;66:39-44.
- 26. Sinaki M, Khosla S, Limburg PJ, Rogers JW, Murtaugh PA. Muscle strength in osteoporotic versus normal women. Osteoporos Int 1993;3:8-12.
Fig. 1Device setting for isometric back extensor strength evaluation with portable dynamometer. (A) Portable dynamometer (Power Track II Commander Muscle Tester; JTECH Medical, Midvale, UT, USA) composed of a compression force sensor (left) and a control/recording unit (right). (B) Square back plate. This wooden plate interface with subjects' back muscle and transmit compression pressure to the attached force sensor unit. (C) Specially designed chair. Vertical linear groove in seatback center is designed for holding a sensor unit with a ruler to set the sensor unit in a specific height. (D) Subject seated in the chair and the pelvis fixed by fastening the seat belt. Square back plate was shown in the red elliptical circle between chair seatback and subjects. The force sensor unit is not visible because it is inserted into the center groove of the chair.
Fig. 2Isometric back extensor strength evaluation with isokinetic dynamometer machine (PrimusRS; BTE Technologies Inc., Hanover, MD, USA).
Fig. 3Correlation between PrimusRS and new method (p<0.001).
Fig. 4Bland-Altman plot for muscle strength data. Dotted line indicates 95% limits of agreement ranging from −63.1 to 61.1 N.
Table 1Demographic and isometric back extensor strength data of study subjects
Table 2Intraclass correlation coefficient (ICC)
Figure & Data
Citations
Citations to this article as recorded by

- Strength and endurance of the lumbar extensor muscles and their predictors: A cross-sectional study in healthy subjects
Daniela Vlazna, Blanka Adamova, Peter Krkoska, Viktoria Kokosova, Katerina Matulova, Tamara Barusova, Michaela Sladeckova
Journal of Electromyography and Kinesiology.2025; 80: 102973. CrossRef - Weak trunk extensors in adult spinal deformity patients are related to sagittal malalignment and kinematic limitations
Maria Saadé, Rami Rachkidi, Ali Rteil, Elma Ayoub, Elena Jaber, Celine Chaaya, Nabil Nassim, Elio Mekhael, Rami Rehayem, Julien Abi Nahed, Bilal Ramadan, Mohamad Karam, Ismat Ghanem, Abir Massaad, Ayman Assi
Gait & Posture.2025; 117: 292. CrossRef - Evaluating pelvic rotator strength: Investigating asymmetry and its correlation with pelvic rotation angle during active straight leg raise
Joo-Young Jeon, Oh-Yun Kwon, Chung-Hwi Yi, Sung-Min Ha, Jun-Hee Kim, Seung-Yoon Han
Journal of Back and Musculoskeletal Rehabilitation.2025; 38(1): 93. CrossRef - Precision and Reliability of a Dynamometer for Trunk Extension Strength and Steadiness Assessment
Franciele Parolini, Márcio Goethel, Johan Robalino, Klaus Becker, Manoela Sousa, Barbara C. Pulcineli, Ulysses F. Ervilha, João Paulo Vilas-Boas, Rubim Santos
Applied Sciences.2025; 15(8): 4081. CrossRef - Reliability of the EN-TreeM dynamometer for measuring back strength in university athletes with chronic low back pain
Sumbul Ansari, Md Farhan Alam, Saurabh Sharma
Sport Sciences for Health.2024; 20(2): 517. CrossRef - The effects of back extensor strength in different body positions on health-related quality of life in patients with degenerative spinal deformity
Sen Yang, Can Chen, Yong Tang, Kai Li, Xueke Yu, Jiulin Tan, Chengmin Zhang, Zhigang Rong, Jianzhong Xu, Fei Luo
Journal of Back and Musculoskeletal Rehabilitation.2024; 37(2): 503. CrossRef - Effects of Wrist Stability Training Combined with Grip Strength Exercise on Pain and Function in Patients with Nonspecific Chronic Wrist Pain
Seung-Ji Hong, Mi-Young Lee, Byoung-Hee Lee
Medicina.2024; 60(7): 1144. CrossRef - Dynamometry for the assessment of trunk muscle strength in postpartum women with pregnancy-related posterior pelvic girdle pain: A reliability study
Fahimeh-Sadat Jafarian, Mahmonir Jafari-Harandi, Gillian Yeowell, Ebrahim Sadeghi-Demneh
International Journal of Osteopathic Medicine.2024; 54: 100738. CrossRef - Reliability and criterion validity of handheld dynamometry for measuring trunk muscle strength in people with and without chronic non-specific low back pain
Shouq Althobaiti, Deborah Falla
Musculoskeletal Science and Practice.2023; 66: 102799. CrossRef - Back Extensor Strength as a Potential Marker of Frailty Using Propensity Score Matching and Machine Learning
Taewook Kim, Gowun Kim, Hee-won Park, Eun Kyoung Kang, Sora Baek
Journal of Clinical Medicine.2023; 12(19): 6156. CrossRef - Trunk muscle dysfunction in patients with myotonic dystrophy type 2 and its contribution to chronic low back pain
Daniela Vlazna, Peter Krkoska, Michaela Sladeckova, Olesja Parmova, Tamara Barusova, Karolina Hrabcova, Stanislav Vohanka, Katerina Matulova, Blanka Adamova
Frontiers in Neurology.2023;[Epub] CrossRef - Effectiveness of a cardiac rehabilitation program on biomechanical, imaging, and physiological biomarkers in elderly patients with heart failure with preserved ejection fraction (HFpEF): FUNNEL + study protocol
Antonio Ignacio Cuesta-Vargas, Iván José Fuentes-Abolafio, Celia García-Conejo, Estíbaliz Díaz-Balboa, Manuel Trinidad-Fernández, Daniel Gutiérrez-Sánchez, Adrián Escriche-Escuder, Lidia Cobos-Palacios, Almudena López-Sampalo, Jose Maria Pérez-Ruíz, Crist
BMC Cardiovascular Disorders.2023;[Epub] CrossRef - Comprehensive Physical Work Capacity Evaluations for Korean Farmers Assessed in Healthy Volunteers
Seungsu Jeong, Gowun Kim, Hee-won Park, Sora Baek
Annals of Rehabilitation Medicine.2023; 47(6): 468. CrossRef - Test-retest reliability of a load cell setup, Ito, and timed loaded standing tests for measuring muscle strength and endurance in older adults with and without hyperkyphosis
Fatemeh Keshavarzi, Fatemeh Azadinia, Saeed Talebian, Omid Rasouli
Musculoskeletal Science and Practice.2022; 58: 102475. CrossRef - Prevalence and Potential Risk Factors for Occupational Low Back Pain Among Male Military Pilots: A Study Based on Questionnaire and Physical Function Assessment
Yizhuo Yang, Shuai Liu, Mengyu Ling, Chaoqun Ye
Frontiers in Public Health.2022;[Epub] CrossRef - Individuals With Impaired Lumbopelvic Control Demonstrate Lumbar Multifidus Muscle Activation Deficit Using Ultrasound Imaging in Conjunction With Electrical Stimulation: A Cross-sectional Study
Panakorn Sungnak, Sranya Songjaroen, Warin Krityakiarana, Hsing-Kuo Wang, Jim Richards, Peemongkon Wattananon
Archives of Physical Medicine and Rehabilitation.2022; 103(10): 1951. CrossRef - Natural aging course of lumbar extensor muscle mass and strength in community-dwelling older women: a 1-year prospective observational study
Dong Hyun Kim, Jinhee Park, Chang Won Lee, Sang Yoon Lee
Aging Clinical and Experimental Research.2022; 34(9): 2099. CrossRef - Practicable performance-based outcome measures of trunk muscle strength and their measurement properties: A systematic review and narrative synthesis
Shouq Althobaiti, Alison Rushton, Ahmad Aldahas, Deborah Falla, Nicola R. Heneghan, Fatih Özden
PLOS ONE.2022; 17(6): e0270101. CrossRef - Prognostic Value of Leg Muscle Strength in Acute Heart Failure Syndrome
JONG-CHAN YOUN, SUK-WON CHOI, HYE SUN LEE, SEONGWOO HAN, EUI-CHEOL SHIN, SANG HONG BAEK, SEOK-MIn KANG
Medicine & Science in Sports & Exercise.2021; 53(1): 19. CrossRef - Assessment of Lumbar Extensor Muscles in the Context of Trunk Function, a Pilot Study in Healthy Individuals
Daniela Vlažná, Peter Krkoška, Matyáš Kuhn, Filip Dosbaba, Ladislav Batalik, Eva Vlčková, Stanislav Voháňka, Blanka Adamová
Applied Sciences.2021; 11(20): 9518. CrossRef - El fenómeno de flexión-relajación lumbar como test diagnóstico en la evaluación de un déficit lumbar. Sensibilidad y especificidad
J. García Díaz, J. Vargas Montes, M.E. Romero Díez
Rehabilitación.2020; 54(3): 162. CrossRef - A mobile delivered self-exercise program for female farmers
Sora Baek, Gowun Kim, Hee-won Park
Medicine.2020; 99(52): e23624. CrossRef - Prediction of Occupational Physical Activities using Inertial Measurement Units and Deep Learning Models
Yishu Yan, Hao Fan, Yibin Li, Elias Hoeglinger, Alexander Wiesinger, Alan Barr, Carisa Harris Adamson
Proceedings of the Human Factors and Ergonomics Society Annual Meeting.2020; 64(1): 879. CrossRef - Comparison of STarT Back Screening Tool and Simmonds Physical Performance Based Test Battery in Prediction of Disability Risks Among Patients with Chronic Low-Back Pain
Chidozie Emmanuel Mbada (PhD PT), Aanuoluwapo Deborah Afolabi (MSc PT), Olubusola Esther Johnson (PhD PT), Adesola Christianah Odole (PhD PT), Taofik Oluwasegun Afolabi (MSc PT), Odunayo Theresa Akinola (PhD PT), Moses Oluwatosin Makindes (BMR PT)
Rehabilitacja Medyczna.2019; 22(4): 4. CrossRef - Wearable Monitoring Devices for Biomechanical Risk Assessment at Work: Current Status and Future Challenges—A Systematic Review
Alberto Ranavolo, Francesco Draicchio, Tiwana Varrecchia, Alessio Silvetti, Sergio Iavicoli
International Journal of Environmental Research and Public Health.2018; 15(9): 2001. CrossRef