An Orthosis Designed to Control Contracture of the Shoulder and Forearm with Spasticity.

Lee, Ju Kang , Yim, Yoon Myung , Lim, Oh Kyung , Kim, Do Hoon , Oh, Seung Gyun , Kim, Jung Tae
Journal of the Korean Academy of Rehabilitation Medicine 2001;25(6):1069-1073.
1Department of Physical Medicine, Gachon Medical School, Korea.
2Department of Rehabilitation, Gachon Medical School, Korea.
  • 1,815 Views
  • 12 Download
  • 0 Crossref
  • 0 Scopus
prev next

Objective: To develop and evaluate newly designed orthosis to control contracture of the shoulder and the forearm in the spastic upper limb.

Method: Subjects were 6 hemiplegic patients and 1 tetraplegic patient who showed the features of spasticity. Volar wrist hand orthosis was modified to attach a plastic stick which can be easily separated. The orthosis was applied in supinated position for 8 weeks. The shoulder and elbow range of motion was measured every 2 weeks for 8 weeks.

Results: The subjects with spasticity showed significant improvement in the shoulder abduction (p<0.05), adduction (p<0.05) and external rotation (p<0.05) after applying modified volar wrist hand orthosis for 8 weeks. The forearm supination was well maintained for 8 weeks.

Conclusion: The newly designed orthosis showed significant effects in the improvement of the shoulder abduction, adduction and external rotation and the forearm supination of spastic upper limbs.