To investigate the reliability and validity of a new method for isometric back extensor strength measurement using a portable dynamometer.
A chair equipped with a small portable dynamometer was designed (Power Track II Commander Muscle Tester). A total of 15 men (mean age, 34.8±7.5 years) and 15 women (mean age, 33.1±5.5 years) with no current back problems or previous history of back surgery were recruited. Subjects were asked to push the back of the chair while seated, and their isometric back extensor strength was measured by the portable dynamometer. Test-retest reliability was assessed with intraclass correlation coefficient (ICC). For the validity assessment, isometric back extensor strength of all subjects was measured by a widely used physical performance evaluation instrument, BTE PrimusRS system. The limit of agreement (LoA) from the Bland-Altman plot was evaluated between two methods.
The test-retest reliability was excellent (ICC=0.82; 95% confidence interval, 0.65–0.91). The Bland-Altman plots demonstrated acceptable agreement between the two methods: the lower 95% LoA was −63.1 N and the upper 95% LoA was 61.1 N.
This study shows that isometric back extensor strength measurement using a portable dynamometer has good reliability and validity.
Citations
To evaluate the reliability and validity of Korean version of AST (K-AST) as a bedside screening test of apraxia in patients with stroke for early and reliable detection.
AST was translated into Korean, and the translated version received authorization from the author of AST. The performances of K-AST in 26 patients (21 males, 5 females; mean age 65.42±17.31 years) with stroke (23 ischemic, 3 hemorrhagic) were videotaped. To test the reliability and validity of K-AST, the recorded performances were assessed by two physiatrists and two occupational therapists twice at a 1-week interval. The patient performances at admission in Korean version of Mini-Mental State Examination (K-MMSE), self-care and transfer categories of Functional Independence Measure (FIM), and motor praxis area of Loewenstein Occupational Therapy Cognitive Assessment, the second edition (LOTCA-II) were also evaluated. Scores of motor praxis area of LOTCA-II was used to assess the validity of K-AST.
Inter-rater reliabilities were 0.983 (p<0.001) at the first assessment and 0.982 (p<0.001) at the second assessment. For intra-rater (test-retest) reliabilities, the values of four raters were 0.978 (p<0.001), 0.957 (p<0.001), 0.987 (p<0.001), and 0.977 (p<0.001). K-AST showed significant correlation (r=0.758, p<0.001) with motor praxis area of LOTCA-II test. K-AST also showed positive correlations with the total FIM score (r=0.694, p<0.001), the selfcare category of FIM (r=0.705, p<0.001) and the transfer category of FIM (r=653, p<0.001).
K-AST is a reliable and valid test for bedside screening of apraxia.
Citations
To evaluate the accuracy of the swallowing kinematic analysis.
To evaluate the accuracy at various velocities of movement, we developed an instrumental model of linear and rotational movement, representing the physiologic movement of the hyoid and epiglottis, respectively. A still image of 8 objects was also used for measuring the length of the objects as a basic screening, and 18 movie files of the instrumental model, taken from videofluoroscopy with different velocities. The images and movie files were digitized and analyzed by an experienced examiner, who was blinded to the study.
The Pearson correlation coefficients between the measured and instrumental reference values were over 0.99 (p<0.001) for all of the analyses. Bland-Altman plots showed narrow ranges of the 95% confidence interval of agreement between the measured and reference values as follows: 0.14 to 0.94 mm for distances in a still image, -0.14 to 1.09 mm/s for linear velocities, and -1.02 to 3.81 degree/s for angular velocities.
Our findings demonstrate that the distance and velocity measurements obtained by swallowing kinematic analysis are highly valid in a wide range of movement velocity.
Citations
To assess the intra-rater and inter-rater reliability for measuring tibial torsion measurements by a radiographic method using three-dimensional computed tomography reconstruction (3D-CT) and to compare the physical measures to those of 3D-CT.
The study included 33 children who presented with intoeing gait. Tibial torsion was measured by 3D-CT. Distal reference point was the bimalleolar axis. Proximal reference points were the transtibial axis and posterior condylar axis. Physical measurements included thigh-foot angle (TFA) and bimalleolar angle (BMA). 3D-CT measurement and physical measurement were performed twice at both lower extremities by each rater. The intra-rater and inter-rater reliability were calculated by intraclass correlation coefficiency (ICC). The relationship between radiological and physical examination was calculated by Spearman correlation coefficient.
The 3D-CT measures for tibial torsion were reliable within individual raters and between different raters. However, physical measures for tibial torsion were reliable within an individual rater but not reliable between raters. The 3D-CT measures by any proximal reference axis were more reliable within a rater and between raters than physical measurements. There was no significant impact introduced by the selection of the proximal reference axis. The correlation coefficiency between 3D-CT and physical measurement methods was low.
Because the 3D-CT measurements for tibial torsion are more reliable than physical measurements, we recommend that accurate diagnosis of internal tibial torsion should be detected by using 3D-CT measurements. Also, considering the disadvantages of radiological measurements, physical measurement may be used for short term follow-up by same raters, as intra-rater reliability is relatively good.
Citations
To investigate the inter-rater agreement for the clinical dysphagia scale (CDS).
Sixty-seven subjects scheduled to participate in a video-fluoroscopic swallowing study (VFSS) were pre-examined by two raters independently within a 24-hour interval. Each item and the total score were compared between the raters. In addition, we investigated whether subtraction of items showing low agreement or modification of rating methods could enhance inter-rater agreement without significant compromise of validity.
Inter-rater agreement was excellent for the total score (intraclass correlation coefficient (ICC): 0.886). Four items (lip sealing, chewing and mastication, laryngeal elevation, and reflex coughing) did not show excellent agreement (ICC: 0.696, 0.377, 0.446, and κ: 0.723, respectively). However, subtraction of each item either compromised validity, or did not improve agreement. When redefining 'history of aspiration' and 'lesion location' items, the inter-rater agreement (ICC: 0.912, 0.888, respectively) and correlation with new videofluoroscopic dysphagia score (PCC: 0.576, 0.577, respectively) were enhanced. The CDS showed better agreement and validity in stroke patients compared to non-stroke patients (ICC: 0.917 vs 0.835, PCC: 0.663 vs 0.414).
The clinical dysphagia scale is a reliable bedside swallowing test. We can improve inter-rater agreement and validity by refining the 'history of aspiration' and 'lesion location' item.